| Series | LOF400-20BxxR2 |
| Model | LOF400-20B24R2 |
| Type | Medical Power Supplies |
| Brand | MORNSUN |
| Input Voltage Range (Rated) | 100–240 VAC |
| Input Voltage Range (AC) | 80–264 |
| Input Voltage Range (DC) | 110–370 VDC |
| Input Voltage Frequency | Hz; AC input: 47–63 Hz; Rated input (Certified voltage): 5.0 Hz |
| Input Current | 115VAC: 5.0A max; 230VAC: 2.5A max; 115VAC: 40A |
| Inrush Current | 230VAC: 80A; 115VAC: 0.98A (Cold start) |
| Power Factor | 0.95 @230VAC |
| Input Fuse | 8A (Built-in fuse) |
| Hot Plug | Unavailable |
| Output Voltage Accuracy | ±3.0% (Full load range) |
| Line Regulation | ±0.5% (Rated load) |
| Load Regulation | ±1.0% (0% - 100% load) |
| Minimum Load | 0 (At room temperature, 230VAC input (when PS_ON is at; high potential)) |
| Stand-by Power Consumption | Vdd voltage: 3.3-15V, Vdd supply current requirement:: 0.5W; 5mA.When PS_ON is at a high level to SGND, the power: W; output is turned off: W; 12V: 120W; W (20MHz bandwidth) |
| Ripple & Noise | 24V: 200mV max; (peak-peak value): mV; 48V: 250mV max |
| Temperature Coefficient | ±0.03%/℃; Air cooling (250W Load): 30%/℃ |
| Hold-up Time | 230VAC, rated load, 25℃: ms; 23CFM (400W Load): 16 ms |
| Short Circuit Protection | Recovery time <3s after the short circuit disappear.: Hiccup, continuous, self-recover; ≥105% Io, hiccup, self-recover (Normal temperature,; high temperature) |
| Over-current Protection | 100-240VAC, rated load: hiccup,; ≥105% full load after derating, hiccup,; Low temperature: hiccup,; self-recover hiccup,; 12V: ≤16.0VDC (Hiccup, self-recover) hiccup, |
| Over-voltage Protection | 24V: ≤33VDC (Hiccup, self-recover); 48V: ≤60VDC (Hiccup, self-recover); 115VAC, 100% load: 65–90 %Vo (Over-temperature; protection start) |
| Over-temperature Protection | Recommended external: ℃; Over-temperature: ℃; wind environment: 23CFM: 50 to 75℃; protection release: ℃; Provide 12V/0.5A output for the fan, with; voltage accuracy of-15% to +15% under t |
| Fan power | he premise that the main output is greate; r than 20% of the rated load (12V) |
| 2. *The “Tip and barrel method” is | and 0.1uF ceramic capacitor, please refer to AC-DC (used for ripple and noise test, output parallel 10uF electrolytic capacitor) |
| Isolation (Input–Output) | 4000 VAC |
| Isolation | 2000 VAC (Electric strength test for 1min., leakage current <5mA) |
| Isolation (Output–⏚) | 1500 VAC |
| Isolation (Input–⏚) | 100 VAC |
| Insulation Resistance | Input - output Relative humidity: < 95%RH, no condensation: 100 MΩ; Test voltage: 500VDC: MΩ; Output -: 100 MΩ |
| Operating Temperature | -40 to 85℃ |
| Storage Temperature | -40–85℃ |
| Storage Humidity | 10–95 %RH; No condensation: %RH |
| Operating Humidity | 20–90%/℃; 23CFM -40℃ to +50℃: 0 %/℃; rated power 400W +50℃ to +70℃: 1.87 %/℃; +70℃ to +85℃: 3.0 %/℃; Operating 230VAC: %/℃; -40℃ to +45℃: 0 %/℃; temperatu: %/℃; Air cooling: %/℃; re derating +45℃ to +60℃: 1.5 %/℃ |
| Power Derating | rated power: %/VAC; 250W +60℃ to +70℃: 1.0 %/VAC; 115VAC -40℃ to +40℃: 0 %/VAC; +40℃ to +70℃: 1.42 %/VAC; Input voltage 80VAC-115VAC: 1.142 %/VAC; derating 115VAC-264VAC: 0 %/VAC; Touch current: <0.1mA %/VAC |
| Leakage Current | Earth leakage current: <0.5mA max; BS EN62368-1, EN62368-1(Report); Design refer to IEC/UL62368-1, (240VAC, 60Hz) |
| Safety Standard | GB4943.1; –IEC/EN60335-1; IEC/EN61558-1; –IEC/EN/ES60601-1; CLASS I (with PE and must be |
| Safety Class | connected)/; CLASS II (without PE) |
| MTBF | ≥200,000 h (MIL-HDBK-217F@25℃) |
| Warranty | 3 years (Ambient temperature: <25℃) |
| Item | Operating Conditions Min.: Typ. typ, Max. max Unit; External power: Unit; Vdd voltage 3.3: 15 Unit; supply: Unit |
| PS_ON Input Signal | Power on PS_ON low 0: 0.6V; Power off PS_ON high 3.3: 24V; High level High 2: 5V; Low level Low 0: 1V; The PG signal goes high: V |
| PG Signal | Power on with 10ms to 500ms delay 10: 500 ms; after power set up: ms; The TTL signal goes low at: ms; Power: ms; least 1ms before output 1: ms; off/Power fail: ms; below 90% of rated value: ms; The total compensation voltage value when Vs+ and Vs-: ms; (pins S1 and S2 of CN2) are respectively short-circuited: ms |
| Remote compensation | --: 0.5V; to the output load terminals (RS+ short-circuited to +Vo,: V; RS- short-circuited to -VO): V; Main output Normal output: Green on V |
| LED Signal | Power off (AC without Input): Light off; When: RS+ and RS- are connected to the system, (status; indication) |
| Remote Sense | with: function of remote voltage compensation,; if not needed, left RS+ and RS- open |
| High and Low Temperature | GB2423.1, typ, IEC60068-2-1 max (Working +85℃, -40℃) |
| Sinusoidal Vibration | GB2423.10, typ, IEC60068-2-6 max (10 - 500Hz, 2g, three directions of X, Y, Z axis) |
| Alternating Hot and Humid | GB2423.4, typ, IEC60068-2-30 max (+25℃, 95%RH - +60℃, 95%RH) |
| Low Temperature Storage | GB2423.1, typ, IEC60068-2-1 max (-40℃) |
| High Temperature Storage | GB2423.2, typ, IEC60068-2-2 max (+85℃) |
| High Temperature Aging | GB2423.2, typ, IEC60068-2-2 max (+50℃ (23CFM, Full load)) |
| Normal Temperature Aging | GB2423.1, typ, IEC60068-2-1 max (+25℃ (23CFM, Full load)) |
| Temperature Shock | GB2423.22, typ, IEC60068-2-14 max (-40℃ to +85℃) |
| Temperature Cycle | GB2423.22, typ, IEC60068-2-14 max (-25℃ to +50℃) |
| Hot and Humid | GB2423.50, typ, IEC60068-2-67 max (+85℃, 85%RH) |
| High Temperature Elevation | GB2423.26, typ, IEC60068-2-41 max (+60℃, 54KPa) |
| Low Temperature Elevation | GB2423.25, typ, IEC60068-2-40 max (-25℃, 54KPa) |
| Constant Humid and Hot | GB2423.3, typ, IEC60068-2-78 max (+40℃, 95%RH) |
| Sinusoidal Vibration Response | GB/T typ, 11287-2000, max IEC60255-21-1 (10 - 150Hz, 1g, three directions of X, Y, Z axis) |
| Sinusoidal Impulse Response | 15g, pulse duration 11ms, three times in each direction: of X, IEC60255-21-2; GB/T typ, 114537-1993, max IEC60255-21-2 |
| Packaging Drop | GB2423.8, typ, IEC68-2-32 max (1m, one corner, three edges and six sides) |
| Case Material | Open frame |
| Dimension | 127.00mm × 76.20mm × 27.50mm |
| Weight | 410g (Typ.) |
| Cooling Method * | Air cooling (250W) / 23CFM (400W) |
| EMC Emissions | CE: CISPR32/EN55032 (CLASS I equipment, CLASS B; CLASS II (equipment, CLASS A)) |
| EMC Immunity | Harmonic current: IEC/EN61000-3-2 CLASS A and CLASS D; ESD: IEC/EN 61000-4-2 Contact ±8KV/Air ±15KV (perf. Criteria A); RS: IEC/EN 61000-4-3 10V/m (perf. Criteria A); EFT: IEC/EN 61000-4-4 ±2KV (perf. Criteria A); Surge: IEC/EN 61000-4-5 ±2KV/±4KV (perf. Criteria A); CS: IEC/EN61000-4-6 10 Vr.m.s (perf. Criteria A); Voltage drop: 40% 240 ,10/12 Period(50/60Hz) (perf. Criteria B); Voltage interruption: 0% 240 ,250/300 periods (50/60Hz) (perf. Criteria C) |